카테고리 보관물: OLYMPUS

OLYMPUS

CONFOCAL MICROSCOPE

Posted on 

Next Generation FLUOVIEW for the Next Revolutions in Science

OLYMPUS CONFOCAL FV3000

The FLUOVIEW FV3000 series of confocal laser scanning microscopes meets some of the most difficult challenges in modern science. Featuring the high sensitivity and speed required for live cell imaging as well as deep tissue observation, the FV3000 confocal microscope enables a wide range of imaging modalities, including macro-to-micro imaging, super resolution microscopy, and quantitative data analysis. Choose between upright and inverted microscope frames suited for a range of life science applications, including developmental biology, stem cell research, electrophysiology, cancer research, slide imaging, and more.

Main Laser

405 nm: 50 mW, 488 nm: 20 mW, 561 nm: 20 mW, 640 nm: 40 mW 

One optional laser port for the sub laser combiner or optional laser unit

Optional Laser

445 nm: 75 mW, 514 nm: 40 mW, 594 nm: 20 mW

Sub laser combiner Laser as follows (max. 3 laser units )
or direct connection to main combiner (single laser unit)

Scanner

FV3000FV3000RS
2 silver-coated galvanometer scanning mirrors2 silver-coated galvanometer scanning mirrors
1 silver-coated resonant and
1 silver-coated galvanometer scanning mirrors

Scanning Method

Scanning Resolution: 64 × 64 to 4096 × 4096 pixels
Scanning Speed (One Way): 512 × 512 with 1.1 s – 264 s. pixel time : 2 μs – 1000 μs
Scanning Speed (Round Trip): 512 × 512 with 63 ms – 250 ms / 256 × 256 with 16 ms – 125 ms
Optical Zoom: 1X – 50X in 0.01X increments
Scan Rotation: Free rotation (360 degrees) in steps of 0.1 degree
Scanning Mode: PT, XT, XZ, XY, XZT, XYT, XYZ, XYλ, XYZT, XYλT, XYλZ, XYλZT
ROI scanning, rectangle clip, ellipse, polygon, free area, line, free line, and point

;tornado mode only for stimulation

Galvanometer Scanner (Normal Imaging)

FV3000FV3000RS
Scanning Resolution: 512 × 32 to 512 × 512 pixels
Scanning Speed: 30 fps at 512 x 512, 438 fps at 512 × 32
Optical Zoom: 1X – 8X in 0.01X increments
Scanning Mode: XT, XZ, XY, XZT, XYT, XYZ, XYλ, XYZT, XYλT, XYZ, XYλZT
ROI Scanning, Rectangle Clip, Line

Resonant Scanner (High-Speed Imaging)

hIGH sENSITIVITY – SPECTRAL DETECTOR

Cooled GaAsP photomultiplier, 2 channels
Motorized Volume Phase Holographic transmission diffraction grating, motorized adjustable slit
selectable wavelength bandwidth: 1–100 nm, wavelength resolution: 2 nm

Detector Module

SPECTRAL DETECTOR

Multi-Alkali photomultiplier, 2 channels
Motorized Volume Phase Holographic transmission diffraction grating, motorized adjustable slit,
selectable wavelength bandwidth: 1–100 nm, wavelength resolution: 2 nm

Detector Module

FV3000 광로도
FV3000 광로도